Autoclave Validation

 


1.1  Verification and sequential operation of sterilization cycles

1.1.1        Test procedure: The prime objective of this test is to verify the physical qualification and integrity of the Autoclave. A minimum of three observations to be made.

1.1.2  Vacuum leak test

1.1.2.1             Objective : To check the sequential operation of vacuum leak test cycle

1.1.2.2             Procedure :

·       Ensure that all the critical instruments are in a calibrated state.

·         Select the vacuum leak test cycle from HMI as per SOP.

·         During vacuum leak test, the equipment should operate as follows:

·         Vacuum level reaches in the chamber and note the reading (P1)  by manual observation in PLC. (Negative pressure should less than -0.85 bars).

·         2nd: Hold the vacuum for 5 minutes i.e. delay before hold (Should not be more than 5 min and less than 2 minutes).

·         3rd:   Note the reading (P2) after 5 minutes delay period.

·         4th:   Continue holding of vacuum for further 10 minutes.

·         5th:   Note the reading (P3) of vacuum after the 10 minutes.

·         Equipment the operator is to note the pressure of P1, P2 and P3 from print out.

·         Calculate the rate of vacuum drop as per the following formula

                          Rate of vacuum drop = 

·         Attach the cycle printout with the identification of all sequential steps.

1.1.2.3              Acceptance criteria: The maximum acceptable leak should be less than 10mm of Hg/ 10 minutes of hold.