Qualification/ Re-Qualification of Autoclaves


1.0             
OBJECTIVE:


 

1-1  To lay down a procedure for Qualification and Re-qualification of steam sterilizers / Autoclaves.

 

2.0              SCOPE:

 

2.1 This procedure is applicable to all Autoclaves performance / re-qualifications performed in Aurobindo Pharma Ltd. Unit-IV.

 

3.0              RESPONSIBILITY:

 

3.1              Validation Group:

Is responsible for preparation of Autoclave Qualification / Re-Qualification plan/ protocols and also responsible for execution of qualification activity as per the approved protocol, with required support from cross functional departments (User, Engineering, and QC).

 

3.2              Engineering / Instrumentation:

Engineering/Instrumentation department is responsible to review the protocol, and support for the execution activity.

 

3.3              User department:

Is responsible to review the protocols, support the qualification activity wherever applicable, and review the qualification data/ activity.

 

3.4              Quality Control:

To review the protocol and perform sample testing, wherever applicable.

 

3.5              Head- QA:

Head-QA is responsible to approve the protocol and report.

 

4.0              ACCOUNTABILITY:

 

4.1                It is the responsibility of respective department heads to ensure the compliance and the concerned personnel performing the activities are trained on this SOP.

 

4.2                Head – QA is accountable for overall compliance of the SOP.

 

5.0              DEFINITION:

 

5.1                Biological indicator: An inoculated microbial carrier contained within its primary pack ready for use.

 

5.2                Cold Spot: The locations with in the load zone that achieves the lowest process lethality (Fo) and / or the lowest distribution temperatures during the sterilization process.

 

5.3                D-value: The exposure time under a defined set of conditions to cause a 1-logarithm reduction in the population of a particular microorganism.

 

5.4                Drain: Drain which is situated at the lowest part of the sterilizer chamber to control the discharge of air/non-condensable gases or air and condensate from the sterilizer chamber.

 

5.5                Equilibration time: The period that elapses between the attainment of the minimum exposure temperature at the reference measure point (typically the drain) and the attainment of the sterilization temperature at all points within the load. This period is an indication of the ability to properly remove air and heat the load items; consequently, it is typically only evaluated by placing HP probes in porous/hard goods loads.

 

5.6                Lag Time: The time difference between the attainment of the minimum sterilization temperature (any point) and the attainment of the sterilization temperature at all points within the load. It is typically only evaluated by placing HP probes in liquid loads (preferably media).

 

5.7                Fo value: The time (in minutes) required to provide the lethality equivalent to that which is provided at 121°c for one minute.

 

5.8                Holding /Hold time: Period for which the temperature of all points within the sterilizer is held within the sterilization temperature band. (NOTE. The holding time follows immediately after the equilibration time. The extent of the holding time is related to the sterilization temperature).

 

5.9                Overkill: A cycle which provides more than a 12-log reduction of a resistant biological indicator with a known D-value of not less than 1 minute. This approach assures substantially greater than a 12-log reduction of the bioburden and therefore only minimal information on the bioburden is required.

 

5.10            Porous Load: Material in an autoclave which is sterilized by heat transfer from condensation of steam directly on to its contaminated (or potentially contaminated) surfaces.

 

5.11            Probability of a Non-Sterile Unit (PNSU): The number expressing the probability of finding a non-sterile unit in a known number of sterilized units. For sterilization processes in the healthcare industry a maximum probability of 1 non-sterile unit in a million units is usual.

 

5.12            Sterility Assurance Level (SAL): An estimate of the effectiveness of a sterilization process generally developed through detailed knowledge of the process itself and the organisms likely to be present.

 

5.13            Sterilization: Process undertaken to render an object(s) sterile.

 

5.14            Sterilization cycle : Automatic sequence of operating stages performed in a sterilizer for the purpose of sterilization

 

5.15            Sterilization temperature range: Range of temperatures, expressed as the sterilization temperature and the maximum allowable temperature which may prevail throughout the load during the holding time.

 

6.0              FLOW CHART :

 

6.1        Not Applicable.

 

7.0              PROCEDURE:

 

7.1         Process development:

7.1.1           Process development studies will be performed to establish new parameters & new load configuration and will be recorded in the format provided as an Annexure-1.

7.1.2           Qualification Parameters shall be fixed based on the out come of the trail data and shall be incorporated in to respective SOPs to be followed for further qualification studies and / or routine usage.

7.1.3           Process/cycle development reports (Annexure-1) shall be archived along with qualification documents in QA archival. These reports shall be the basis to prepare user SOPs and qualification protocols etc.

7.2   Protocol Preparation And Approval:

7.2.1           Autoclave qualification Plans / protocols shall be prepared as per the SOP QA076 (Performance qualification / Re-qualification).

7.2.2           Protocol should contain the details about all the tests and acceptance criteria for that particular qualification / re-qualification.

7.2.3           Performance Qualification / Re-qualification execution shall be carried out by responsible cross functional teams of user, engineering and QA.

7.2.4           Inclusion of any new load shall be qualified, through respective protocol or with an addendum in case of independent activity. This load shall be a part of plan & protocol for subsequent routine requalification based on the process developments, as applicable.

7.2.5           All the applicable tests described in this SOP shall be included in the protocol.

7.3   Prerequisites to start execution:

7.3.1           Note down all the prerequisites checks and other checks to be performed as described in the respective templates but not limited to, along with the defined and measurable specific acceptance criteria.

7.3.2           Calibration status of the all temperature and pressure devices of the respective steam sterilizer/Autoclaves should be reviewed and recorded before commencement of performance qualification/re-qualification.

7.3.3           Ensure that Data logger and temperature sensors used in the qualification activity, shall be calibrated and with in the due period, prior to start of activity.

7.3.4           All other operational attributes & SOPs identified in the protocol shall be in place prior to start of the activity.

 

7.4 Verification of Vacuum Leak Test:

7.3.5           Check and ensure that the vacuum leak test parameters are as mentioned in the respective operation SOP before start the cycle.

7.3.6           The prime objective of this test is to validate the integrity of the autoclave chamber.

7.3.7           If vacuum leak test fails, rectify leakage and perform vacuum leak test again until it passes.

7.3.8           The maximum acceptable leak should be less than 1.3 mbar/minute or 0.013 bar over a period of 10 minutes or as per design specifications.

7.5 Verification of Bowie-Dick Test:

7.3.9           The Bowie-Dick Test (air removal test) is to evaluate the efficiency of air removal during the pre-vacuum phases of sterilization cycle. The Bowie-Dick test is a diagnostic test of a sterilizer's ability to remove air from the chamber and prevents residual air remaining in porous masses.

7.3.10       A ready “Bowie and Dick test kit” or “Bowie and Dick test sheet” should be used.

7.3.11       Check and ensure the validity of the test kit / sheet to be use with respect to its expiry date.

7.3.12       Place the “Bowie-Dick test kit” near to drain at a distance of above 5 cm.

7.3.13       Select the Bowie and Dick test cycle in PLC and perform the test as described in the respective standard operating procedure of the equipment.

7.3.14       The Bowie-Dick cycle shall be run for a hold period of 3.5 minutes at 134-138°C unless other conditions for e.g. 17 minutes at 121°C or specifically recommended by suppliers of proprietary kits.

7.3.15       After completion of the cycle, retrieve and examine the test sheet for the compliance against the acceptance criteria (Uniform color change should observe on Bowie-dick indicator).

7.3.16       Document the test details and attach the test sheet to the protocol.

 

7.6 Empty Chamber Heat Distribution Studies:

7.3.17       Empty Chamber Heat Distribution studies shall be done for new autoclaves or for autoclaves moved from one location to another or if steam supplies are significantly changed studies and no need to be perform during the periodic re-qualification of the steam sterilizers/Autoclaves.

7.3.18       Each Empty Chamber Heat Distribution trial shall be done using at least 12 sensors or shall be decided based on the size / volume of the equipment

7.3.19       One shall be located in the drain

7.3.20       One shall be located adjacent to the control or monitoring sensor within the chamber

7.3.21       One shall be located at the geometric dead centre of the chamber

7.3.22       The remain shall be distributed throughout the chamber

7.3.23        

7.3.24       Place the data logger, pre calibrated temperature sensors (identified with numbers) inside sterilizer chamber, as per the temperature sensor location diagram enclosed for the respective protocol by covering all locations of the sterilization chamber.

7.3.25       Select the sterilization cycle and verify the sterilization cycle parameters with respective standard operating procedure of the sterilizer/Autoclave.

7.3.26       Carry out the empty chamber heat distribution studies with temperature logging interval of not more than 5 seconds.

7.3.27       After completion of sterilization cycle stop the data logger and open the non sterile door of the sterilizer/Autoclave.

7.3.28       Take and check out the Temperature profile from the Data logger, equipment temperature printout and strip chart recorder of machine and attach with the executed protocol.

7.3.29       Calculate the maximum, minimum, average temperatures within the probes during sterilization hold period of Empty Chamber Heat Distribution cycle.

7.3.30       Photographs showing placement of sensors should be included in documentation.

7.3.31       Three consecutive successful runs of empty chamber heat distribution shall be performed during initial performance qualification to confirm uniform heat distribution in the sterilizer chamber by mapping the temperatures at different locations and to identify the cold spot (if any) at which the lowest Fo- value has accumulated with in the chamber throughout end of the hold period.

7.3.32       If more than one sensor fails (except sensor located at drain point) during the study, shall be considered invalid and may be repeated.

7.3.33       Empty chamber runs are not generally done with a completely empty chamber. It is permitted to have a “rack” within the chamber to support the sensors.

7.3.34        

7.3.35       Check the calibration reports for its compliance with respect to the acceptance criteria and attach the Calibration reports along with executed protocol.

7.3.36       The F0 - value of all the temperature sensors of data logger should not be less than of sterilization hold time.

7.3.37       The temperature measured, throughout the sterilization holding time, should be within the 121.0 to 124.0 °C.

7.3.38       Throughout the dwell time, the temperature variation of each sensor does not fluctuate by more than 20C.

7.3.39       Throughout the dwell time at that particular time the temperature variation of all sensors does not fluctuate by more than 20C.

 

7.4 Loaded Chamber Heat Distribution Studies:

7.3.40       Loaded Chamber Heat Distribution studies shall be done for new Autoclaves or for autoclaves moved from one location to another or if steam supplies are significantly changed, if any change in the load pattern, any addition of new loads. And no need to perform during the periodic re-qualification of the Autoclaves.

7.3.41       Each Loaded Chamber Heat Distribution trial shall be done using at least 12 sensors or shall be decided based on the size / volume of the equipment

7.3.42       One shall be located in the drain

7.3.43       One shall be located adjacent to the control or monitoring sensor within the chamber

7.3.44       One shall be located at the geometric dead centre of the load

7.3.45       The remain shall be distributed throughout the load

7.3.46       Sensors shall be placed beside (not within) load items

7.3.47        

7.3.48       Place the data logger temperature sensors (identified with numbers) inside the sterilizer chamber beside (not within) the load, as per the temperature sensor location diagram enclosed in authorized protocol covering of all critical locations in the sterilization chamber.

7.3.49       Select the sterilization cycle and verify the sterilization cycle parameters with respective standard operating procedure of the sterilizer/Autoclave.

7.3.50       Carry out the loaded chamber heat distribution studies with temperature logging interval of not more than 5 seconds.

7.3.51       After completion of sterilization cycle stop the data logger and open the non sterile door of the sterilizer/Autoclave.

7.3.52       Take and check out the Temperature profile from the Data logger, equipment temperature printout and strip chart recorder of machine and attach with the executed protocol.

7.3.53       Calculate the maximum, minimum, average temperatures within the probes during sterilization hold period of Loaded Chamber Heat Distribution cycle

7.3.54       Photographs showing the loads and placement of sensors should be included in documentation. 

7.3.55       Three consecutive successful runs of loaded chamber heat distribution with maximum load in its defined configuration shall be performed during initial performance qualification to confirm uniform heat distribution in the sterilizer chamber by mapping the temperatures at different locations and to identify the cold spot (if any) at which the lowest Fo- value has accumulated to the end of the hold period.

7.3.56       The “cold spot” location obtained from Loaded Chamber Heat Distribution Trials shall be compared to the findings obtained from Empty Chamber Heat Distribution Trials.

7.3.57       If more than one sensor fails (except sensor located at drain point) during the study, shall be considered invalid and may be repeated.

7.3.58       Check the calibration reports for its compliance with respect to the acceptance criteria and attach the Calibration reports along with executed protocol.

7.3.59       The Fo - value of all the temperature sensors of data logger should not be less than of sterilization hold time.

7.3.60       The temperature measured, throughout the sterilization holding time, should be within the 121.0 to 124.0 °C.

7.3.61       Throughout the dwell time, the temperature variation of each sensor does not fluctuate by more than 20C.

7.3.62       Throughout the dwell time at that particular time the temperature variation of all sensors does not fluctuate by more than 20C.

 

7.1Loaded Chamber Heat Penetration with Biological indicator Studies:

7.3.63       These studies shall be done for new autoclaves or for Autoclaves moved from one location to another or if steam supplies are significantly changed or for the introduction of new items or new load configurations.

7.3.64       These studies shall be carried out for all types of load configurations and load patterns individually.

7.3.65       Each trial shall be done using at least 12 sensors or shall be decided based on the size / volume of the equipment.

7.3.66       One shall be located in the drain

7.3.67       One shall be located adjacent to the control or monitoring sensor within the chamber

7.3.68       The remain shall be placed within items making up the load and shall be distributed throughout the load.

7.3.69       If a “cold spot” has been identified in previous trials of the autoclave, sensors shall be located at or close to any “cold spot” which may have been identified.

7.3.70        

7.3.71       Place the data logger temperature sensors (identified with numbers) inside the load, as per the temperature sensor location diagram enclosed for the respective protocol.

7.3.72       Each sensor shall be placed within items at locations from which air may be most difficult to remove and to which steam may be most difficult to penetrate.

7.3.73       Place pre enumerated biological indicators (20 numbers) of Geobacillus stearothermophilus (carrier shall be paper strips having minimum106 spores per strip with D121-values of between 1.5 and 4.0 minutes) with properly marked details in the middle of the sterilization load near each temperature sensor kept to check heat penetration, and also at the locations where sensor can not be placed but the location is defined as difficult for steam penetration (if any).

7.3.74       Note: Biological indicators shall be placed only in the third run of each loaded heat penetration study.

7.3.75       In case of rubber stoppers load instead of biological indicator spore strips, spiked rubber stoppers (20 numbers) shall be used. Spiked rubber stopper shall be kept in tyvek or parchment paper to expose in autoclave.

7.3.76       In case of fluid containers load (e.g. media loads in QC sterilizers/Autoclave) instead of biological indicators spore strip, liquid ampoule biological indicators (20 numbers) shall be used.

7.3.77       There shall be a documented justification for the items selected and for the locations within the items selected for BI placement.

7.3.78       Select the sterilization cycle and verify the sterilization cycle parameters with respective standard operating procedure of the sterilizer/Autoclave.

7.3.79       Carry out these studies with temperature logging interval of not more than 5 seconds.

7.3.80       After completion of sterilization cycle stop the data logger, open the non sterile door of the sterilizer/Autoclave and take out the Biological indicators from the load & send to Microbiological Lab for testing as per the respective SOP.

7.3.81       Take and check out the Temperature profile from the Data logger, equipment temperature printout and strip chart recorder of machine and attach with the executed protocol.

7.3.82       Check the equilibration time.

7.3.83       Calculate the maximum, minimum, average temperatures within the probes during sterilization hold period of Loaded Chamber Heat Penetration cycle.

7.3.84       Photographs showing the loads and placement of sensors should be included in documentation.

7.3.85       If more than one sensor fails (except sensor located at drain point) during the study, shall be considered invalid and may be repeated.

7.3.86       Temperature sensors used should be calibrated after completion of activity (post calibration) to ensure that the temperature measurement system is accurate and precise.

7.3.87       Check the calibration reports for its compliance with respect to the acceptance criteria and attach the Calibration reports along with executed protocol. 

7.3.88       At least three (3) successive Heat Penetration trials shall be performed complying with the acceptance criteria for initial qualification by placing the biological indicators in the third run of each defined load configuration.

7.3.89       Consider the lag time for liquid loads and record it.

7.3.90       The F0- value of all the temperature sensor of data logger should be equal to or more than sterilization hold time set (except liquid or nutrient media loads).

7.3.91       For Nutrient media load, F0 - value of all the temperature sensor of data logger should be not less than 15 minutes.

7.3.92       The temperature measured, throughout the sterilization holding time, should be within 121.0 to 124.0 °C.

7.3.93       The equilibrium time should not be more than 30 seconds.

7.3.94       The equilibrium time is not applicable for liquid/ nutrient media loads.

7.3.95       Throughout the dwell time, the temperature variation of each sensor does not fluctuate by more than 20C.

7.3.96       Throughout the dwell time at that particular time the temperature variation of all sensors does not fluctuate by more than 20C.

7.4 Loaded Chamber Heat penetration Studies with biological indicators (Shortened holding period):

7.3.97       Test procedure shall be followed as mentioned in the section 7.8.

7.3.98       Sterilization hold time for all the loads in this study shall be less than sterilization hold time set in the loaded heat penetration studies. This study shall be performed as a one time activity with one run only.

7.3.99       The F0- value of all the temperature sensor of data logger should be equal to or more than sterilization hold time set. (except liquid or nutrient media load)

7.3.100   For Nutrient media load, F0 - value of all the temperature sensor of data logger should be not less than 15 minutes.

7.3.101   The temperature measured, throughout the sterilization holding time, should be within 121.0 to 124.0 °C.

7.3.102   The equilibrium time should not be more than 30 seconds.

7.3.103   The equilibrium time is not applicable for liquid/ nutrient media loads.

7.3.104   Throughout the dwell time, the temperature variation of each sensor does not fluctuate by more than 20C.

7.3.105   Throughout the dwell time at that particular time the temperature variation of all sensors does not fluctuate by more than 20C.

7.3.106   Estimation of Fo Value:

7.3.107   To identify number of equivalent minutes of sterilization at temperature 121.1°C delivered to an item or unit of product calculated using a Z-value of 10°C temperature.

7.3.108   Calculate the Fo value at each temperature mapping probe for each time interval during dwell period by using the following formula:

7.3.109                         F0: ∑ 10 (T-121.1)/z Δ t

7.3.110                         F0 value is time that is equivalent to the sterilization effect at 121.1 ÂºC.

7.3.111                         T = Product temperature at specific time in ºC

7.3.112                         Δ t = Time interval between temperature measurement (in minutes).

7.3.113                          Z = Temperature required to change the D value by a factor of 10. 

7.3.114                          D value is time required to reduce a microbial population by a one-logarithm reduction

7.3.115   Note:  If the Δt is 5 seconds and Z value is 10ºC for sterilization study so the final formula for calculation of Fo values is:

7.3.116                        F0 = ∑10 (T-121.1)/10×0.08

7.3.117   Calculate cumulative Fo value by summing the Fo values for every 5 seconds obtained during sterilization hold period by using the following formula:

7.3.118                         F0 = ∑ (sum of lethal rates)

 

7.4           Re-qualification Programmes:

7.4.1           Re-qualification shall be performed once in six months at minimum.

7.4.2           Heat Penetration and Biological Qualification studies shall be done as part of a time-related re-qualification programme for every autoclave.

7.4.3           One Heat Penetration and Biological Qualification trial shall be done on every defined load (including minimum loads) on each autoclave at least once in 2 years at minimum.

7.4.4           Any modifications in the existing load, incorporation of new loads shall adequately qualify through RQ / Addendum. The number of cycles shall be determined based on the impact assessment performed for respective change control.

7.4.5           A comparison shall be made between requalification data and initial qualification, with suitable comments in summary / conclusion. Any significant variation shall be documented with adequate rationale and proposed actions to prevent potential failures.

7.5           Failure:

7.5.1           Failure to comply with acceptance criteria in any initial qualification study which after investigation results in changes

7.5.2           The engineering or software controls of the autoclave or of the steam supply shall require all qualifications from Empty Chamber Heat Distribution onwards to be repeated.

Ø To defined load configurations or orientation of items within loads shall require all qualifications from Loaded Chamber Heat Distribution onwards to be repeated.

 

7.5.3           Failure to comply with acceptance criteria in re-qualification Heat Penetration or Biological Qualification study shall be handled through the SOP CQA-CP-GEN-014.

7.6           Documentation:

7.6.1           The out come of qualification study with respect to change in load configuration, process parameters etc., shall be incorporated adequately in respective SOPs by user, where ever applicable. Validation group / QA shall ensure this requirement prior to close the executed report.

7.6.2           At the end of each temperature mapping study, data recorded in the data logger temperature mapping storage devise, shall be printed and enclosed as an annexure for executed protocol with sign and date. Required data (For e.g. F0 (in minutes), minimum, maximum and average temperatures) shall be collected from the original print out and recorded in the protocol.

7.6.3           If required data is not available from data loggers print out, original temperature mapping data shall be converted in to MS-EXCEL sheet and F0 (in minutes), minimum, maximum and average temperatures shall be calculated.

7.6.4           Equipment temperature printout and strip chart recorder of machine shall be attached to executed protocol.

7.6.5           Specified Load details, Probe location diagrams & justification and photographs (Empty chamber and loaded chamber) shall be attached as an annexure for executed protocol.

7.6.6           Data logger, temperature sensors (Pre & Post) calibration details / certificates shall be attached to the executed protocol.

7.6.7           Process/Cycle development studies shall be recorded in Annexure-1.

8.0              ABBREVIATION:

 

S. No.

Abbreviation

Full Description

2.

BI’s

Biological Indicators

3.

GPT

Growth Promotion Test

4.

HD

Heat Distribution

5.

HP

Heat Penetration

6.

NA

Not Applicable

7.

NMT

Not More Than

8

PLC

Programme Logic Controller

9.

PNSU

Probability of a Non-Sterile Unit

10.

PQ

Performance Qualification

11.

RQ

Re-Qualification

12.

SAL

Sterility Assurance Level

13.

SOP

Standard Operating Procedure

14.

SP

Sterile process

15.

QA

Quality Assurance

16.

QC

Quality Control

17.

VMP

Validation Master Plan

 

 

9.0              SAFETY, HEALTH AND ENVIRONMENT:

 

·         Identify all the potential hazards during the qualification activity and list adequate measures to be taken, to eliminate the hazards, while developing the qualification protocol

 

10.0          REFERENCE:

 

S. No.

Topic of the SOP

SOP No. (Current Version)

1.

Validation Master Plan

DOC NO.: APL-VMP-02.

2.

Performance qualification / Requalification

SOP QA076-00

3.

Sterilisation of porous loads by moist Heat in Autoclaves

APL-Quality Policy–SP-006

(Version:1.0.0.0)

4.

Numbering and Control of Validation/ Qualification protocols

SOP QA015-06

5.

Good Documentation Practices

SOP QA010-04

6.

Reporting and monitoring of process non conformance in the automated quality management system software

SOP CQA-CP-GEN-014

(Version:2.0.0.0)

·         SHTM 2010

·         EN285 (sterilization by autoclaves).

·         PDA Tech. Report-1. “Steam sterilization”

 

 

11.0          ANNEXURE:

 

·         Annexure-1 - Process/Cycle Development Report.

 

CHANGE HISTORY

Version

Supersedes

Changes made

 

00

 

01

 

 

 

 

 

NA

 

QA081-00

 

 

1.  First issue

 

1. Lag time definition included in the SOP.

2. Process development procedure, format to record the results (Annexure-1) included in the SOP.

3. Side heading verification of calibration status of temperature sensors and data logger changed as prerequisites to start execution.

4. Expanded Requalification programme.

5. Procedure amended vide change request APL-FU4-CC-12-0073.

 

 

 


Previous
Next Post »